Coordinating Multiple PXIe Oscilloscope Modules
8
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module Startup Guide
57
Iteration = 7
Mean = 4031
Mean change = 950
Iteration = 8
Mean = 4031
Mean change = 950
Iteration = 9
Mean = 3551
Mean change = 470
Iteration = 10
Mean = 2271
Mean change = -810
Iteration = 11
Mean = 3717
Mean change = 636
Iteration = 12
Mean = 3717
Mean change = 636
Iteration = 13
Mean = 2270
Mean change = -811
Iteration = 14
Mean = 3873
Mean change = 792
Iteration = 15
Mean = 3078
Mean change = -3
Clock phase reset completed.
Performing clock phase reset for module at slot 6...
Temperature change = 0.25
XML mean = 3117
Target = 3116
Iteration = 1
Mean = 3110
Mean change = -7
Clock phase reset completed.
C:\Program Files (x86)\Keysight\InfiniiVision>
When running the KtScopeCal.exe application with the Sync mode options
"-a", "-p", or "-r", the oscilloscope modules automatically phase-lock their
timebases to the chassis' PXIe 100 MHz reference. Using the PXIe 100 MHz
reference signal is a requirement for Sync mode triggering (see
(Recommended) Phase-Lock Oscilloscope Sampling Clocks"
If you turn off the PXIe 100MHz ref signal on any module, clock phase
alignment between modules is lost, and the red unlocked image will re-appear.
•
When clock phase alignment is lost, use the
KtScopeCal -r
command to
reset/realign the clock phase. For example: