30
DSOX1204A/G Oscilloscopes Service Guide
2
Testing Performance
To verify trigger sensitivity
These tests verify the trigger sensitivity. In these tests, you will apply a sine wave
to the oscilloscope at two bandwidths: 10 MHz and the maximum bandwidth of
the oscilloscope. For each sine wave, you will decrease the amplitude of the signal
to a specified level and check to see if the oscilloscope still triggers.
Table 10
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
10 MHz and 70 MHz, 100 MHz, or
200 MHz sine waves
Keysight N5181A
Power splitter
Outputs differ < 0.15 dB
Keysight 11667A
Power Meter
Keysight N1914A
Power Sensor
Keysight E9304A or N8482A
Cable
BNC, Qty 3
Keysight 10503A
Adapter
N (m) to BNC (f), Qty 3
Keysight 1250-0780
Feedthrough
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Summary of Contents for InfiniiVision DSOX1204A
Page 1: ...Keysight InfiniiVision DSOX1204A G Oscilloscopes Service Guide ...
Page 8: ...8 DSOX1204A G Oscilloscopes Service Guide Figures ...
Page 10: ...10 DSOX1204A G Oscilloscopes Service Guide Tables ...
Page 36: ...36 DSOX1204A G Oscilloscopes Service Guide 2 Testing Performance ...
Page 48: ...48 DSOX1204A G Oscilloscopes Service Guide 5 Replacing Assemblies ...
Page 50: ...50 DSOX1204A G Oscilloscopes Service Guide 6 Replaceable Parts ...
Page 56: ...56 DSOX1204A G Oscilloscopes Service Guide Index ...