Chapter 2: Preparing for Use
To connect optional InfiniiMax oscilloscope probes
27
3
To disconnect the probe, push the small latch on top of the probe connector
to the left, then pull the connector body away from the front panel of the
oscilloscope without twisting it.
Figure 2-8
Disconnecting the Oscilloscope Probe
C A U T I O N
Do not attempt to twist the snap- on probes on or off the oscilloscope’s BNC connector.
Twisting the probe connector body will damage it.
C A U T I O N
For the 8000 series oscilloscopes do not exceed the maximum input voltage rating. The
maximum input voltage for 50
Ω
inputs is 5 Vrms, CAT I. Maximum voltage at 1 M
Ω
impedance is ±100 V (DC + AC) [AC < 10 kHz], CAT I.
!