Chapter 6: Replacing Assemblies
To disconnect and connect Mylar flex cables
101
To disconnect and connect Mylar flex cables
Use this procedure when you are instructed to disconnect or connect a Mylar flex cable.
Such cables and their connectors are fragile; mishandling may damage the cable or
connector.
To disconnect the cable
1
Pry up the retainer slightly at each end of the connector using a small flat-
blade screwdriver. Do not force the retainer; it should remain attached to the
body of the socket.
2
Gently pull the flex cable out of the connector.
Figure 6-4
Disconnecting a Mylar flex cable
To reconnect the cable
1
Ensure that the cable retainer is up, then insert the ribbon cable into the socket,
making sure to observe polarity of the cable with respect to the connector.
2
Push the ends of the retainer down onto the connector body, using a small flat-
blade screwdriver. The retainer should be flush with the connector body when
you are finished.
Figure 6-5
Connecting Mylar flex sables
Summary of Contents for Infiniium 8000 Series
Page 6: ...6...
Page 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Page 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Page 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Page 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Page 158: ...158...
Page 160: ...160 Index...