Using InfiniiMax III+ Series Probes
1
InfiniiMax III+ Series Probes User’s Guide
21
To Ensure Maximum Measurement Accuracy
To increase measurement accuracy, use the oscilloscope’s
Select Probe Head
dialog
box to indicate the type of probe head that is attached to the probe amplifier.
Making this selection allows the oscilloscope to apply the proper type of correction
filter (S parameter) for your measurement case. The correction filter increases
accuracy by flattening the magnitude and phase response of the probe. The
following two sources of S parameters are automatically used for the filter:
• S parameters of the InfiniiMax III+ probe amplifier. These S parameters are
unique to and stored on the probe amplifier.
• S parameters of the specific probe-head model. For example, the
N2836A: DF
Solder-In (Vertical)
selection applies S parameters for the N2836A head held in a
vertical position (perpendicular to the surface of the DUT).
To display the dialog box, click
Probes
>
Channel Setup Menu
.
Figure 14
Select Probe Head Dialog Box
Summary of Contents for InfiniiMax III+ Series
Page 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Page 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Page 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Page 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Page 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Page 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Page 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...