8 Performance Verification for
N7000-Series Probes
Bandwidth Performance Verification
DC Input Resistance Performance Verification
This chapter describes the equipment and procedures needed to verify the
performance of N7000/1/2/3A InfiniiMax III+ probes. The performance measured
in this chapter is of the probe by itself. Keysight high performance real-time
scopes (and sampling scopes under certain conditions) will apply probe correction
that will further enhance the performance of the probes.
NOTE
Due to the very high frequency of the InfiniiMax III+ probing system, it is important to carefully
adhere to the techniques and procedures described in this chapter to accurately measure the
performance.
CAUTION
Electrostatic discharge (ESD) can quickly and imperceptibly damage or destroy high
performance probes, resulting in costly repairs. Always wear a wrist strap when
handling probe components and insure that cables are discharged before being
connected.
Summary of Contents for InfiniiMax III+ Series
Page 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Page 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Page 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Page 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Page 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Page 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Page 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...