Performance Verification for N2830A-Series Probes
7
InfiniiMax III+ Series Probes User’s Guide
135
Performance Test Record
NOTE
The recommended test interval is one year/2000 hours.
Table 25
Test Information
Keysight Technologies
Keysight InfiniiMax III+ Series Probe
Model Number:
Tested by:
Serial Number:
Work Order Number:
Recommended next test date:
Date:
Table 26
Test Results
Test with Probe Heads
(only required to test one)
Test Limits
Result
Pass/Fail
Bandwidth Performance Test
N2836A
≥
13 GHz
DC Input Resistance Performance Test
N2836A
98 to 102 k
W
(differential mode)
49 to 51 k
W
(single-ended mode)
N5441A
98 to 102 k
W
(differential mode)
49 to 51 k
W
(single-ended mode)