Performance Verification for N2830A-Series Probes
7
InfiniiMax III+ Series Probes User’s Guide
119
4
Connect the NMD 3.5 mm (m) to 3.5 mm (f) adapter to the N5477A as shown in
the figure. This adapter is provided with the N5477A.
5
Connect the N5477A to the 1143A power supply and turn on the power supply.
6
Set the 1143A’s probe offset control button to “Zero” so no probe offset is
applied.
7
Connect the N2830A-series probe being tested to the N5442A adapter.
Connect the N5442A to the N5477A adapter as shown in the figure.
8
Connect the probe to the N5443A PV fixture:
a
Insert the probe with ZIF probe head into the PV holder far enough that the
tip wires can easily reach the pinches on the PV fixture.
b
Form the coax cables so that the tip wires are close to the pincher points
before trying to connect the tip wires. The connectors between the probe
head and the probe amp can be rotated to align the probe tip properly to the
punchers. Since the center trace of the PV fixture is above the ground plane,
the probe head should be tipped slightly so the tip wires touch the center
trace and ground plane at the same time.
c
Depress the actuators on the pincher and carefully insert one wire under the
center pincher and the other wire under one of the side pincher. Either
polarity of the probe can be tested and will yield the same results (but
opposite phase) if the probe is working properly.
shows a close up of the tip wires positioned under the pincher.
d
Ideally the probe head should not be angled toward the port 2 side of the PV
fixture, but a slight angle of 5 degrees is acceptable. If angled too much, the
measured BW of the probe will be degraded due to coupling from the trace
to the probe tip.
9
Use the following steps to Install a file to de-embed the adapter (1250-1749)
and the output side of the N5443A. This is the path from the male 2.92 mm
connector to the probe point of the N5443A from Port 1 of the VNA.
a
Create the Touchstone file by cutting and pasting the text in
on page 125. Name the file:
Adaptor_1250_1749___OutputSideOfFixture_N5443A.s2p
.
b
On the VNA, go to menu “Calibration/Fixturing Selections/2 Port
De-embedding” and select Port 1.
c
Set S2P file selection to the file saved in step a.
d
Check the “Enable De-embedding”.
e
Under “Calibration” menu, select “Fixturing ON/off” to turn on
de-embedding.
NOTE
You can also copy the data from the Adobe AIR version of Keysight’s Probe Resource Center
(PRC). Copying this data from the PRC is the simplest most reliable method to get the data. To
download the PRC, visit http://www.Keysight.com/find/PRC.
Summary of Contents for InfiniiMax III+ Series
Page 1: ...Keysight N2830 1 2A and N7000 1 2 3A InfiniiMax III Probes User s Guide...
Page 74: ...74 InfiniiMax III Series Probes User s Guide 2 Using InfiniiMax III Probe Heads...
Page 84: ...84 InfiniiMax III Series Probes User s Guide 3 Maintaining Probe Heads...
Page 92: ...92 InfiniiMax III Series Probes User s Guide 4 Calibration Deskew Procedure...
Page 106: ...106 InfiniiMax III Series Probes User s Guide 5 Specifications and Characteristics...
Page 114: ...114 InfiniiMax III Series Probes User s Guide 6 Performance Plots...
Page 160: ...160 InfiniiMax III Series Probes User s Guide 9 SPICE Models...