Saving calibration data
You must always save new calibration data before cycling instrument power or leaving the calibration state with the
Auto Save feature off. To save calibration data, send CAL:SAVE or save the calibration data from the front panel.
Calibration auto save
The instrument includes a calibration Auto Save feature. This feature automatically saves the calibration data to
non-volatile memory and increments the calibration count when you exit the calibration state.
To enable or disable the CAL auto Save feature, send CAL:ASAV ON or CAL:ASAV OFF. To query the CAL auto Save
state, send CAL:ASAV?
Calibration Test Setup
Front panel connection
Connect all leads to the output terminals as shown in the following figure.
The following figures show the calibration test set-ups. Connect all leads to the input terminal as shown below.
CC Calibration Setup
CV Calibration Setup
32
Keysight EL30000 Series Service Guide