Troubleshooting
Error Messages
Error Messages
If an error condition occurs in the signal generator, it is reported to both the
front panel display error queue and the SCPI (remote interface) error queue.
These two queues are viewed and managed separately; for information on the
SCPI error queue, refer to the
Keysight Signal Generators Programming
Guide
.
NOTE
When there is an unviewed message in the front panel error queue, the
ERR
annunciator appears on the signal generator’s display.
Error Message File
A complete list of error messages is provided in the file
errormessages.pd f
, on
the CDROM supplied with your instrument.
In the error message list, an explanation is generally included with each error
to further clarify its meaning. The error messages are listed numerically. In
cases where there are multiple listings for the same error number, the
messages are in alphabetical order.
Error Message Format
When accessing error messages through the front panel display error queue,
the error numbers, messages and descriptions are displayed on an enumerated
(“1 of N”) basis.
Characteristic
Front Panel Display Error Queue
Capacity (#errors)
30
Overflow Handling
Circular (rotating).
Drops oldest error as new error comes in.
Viewing Entries
Press:
Utility
>
Error Info
>
View Next
(or
Previous
)
Error Page
Clearing the Queue
Press:
Utility
>
Error Info
>
Clear Error Queue(s)
Unresolved Errors
a
Re–reported after queue is cleared.
No Errors
When the queue is empty (every error in the queue has been read, or the queue is cleared), the
following message appears in the queue:
0
No Error Message(s) in Queue
a. Errors that must be resolved. For example, unlock.