Keysight E8357D/67D & E8663D PSG User’s Guide 321
Troubleshooting
Sweep Problems
Sweep Problems
Sweep Appears to be Stalled
The current status of the sweep is indicated as a shaded rectangle in the
progress bar. You can observe the progress bar to determine if the sweep is
progressing. If the sweep appears to have stalled, check the following:
— Have you turned on the sweep by pressing any of the following key
sequences?
Sweep/List
>
Sweep
>
Freq
Sweep/List
>
Sweep
>
Ampl
Sweep/List
>
Sweep
>
Freq & Ampl
— Is the sweep in continuous mode? If the sweep is in single mode, be sure
that you have pressed the
Single Sweep
softkey at least once since
completion of the prior sweep. Try setting the mode to continuous to
determine if the missing single sweep is blocking the sweep.
— Is the signal generator receiving the appropriate sweep trigger? Try setting
the
Sweep Trigger
softkey to Free Run to determine if a missing sweep trigger
is blocking the sweep.
— Is the signal generator receiving the appropriate point trigger? Try setting
the
Point Trigger
softkey to Free Run to determine if a missing point trigger
is blocking the sweep.
— Is the dwell time appropriate? Try setting the dwell time to one second to
determine if the dwell time was set to a value which was too slow or too fast
for you to see.
— Do you have at least two points in your step sweep or list sweep?
Cannot Turn Off Sweep Mode
Press
Sweep/List
>
Sweep
>
Off
.
In the sweep mode menu you can choose to set the sweep to various sweep
types or to turn sweep off.
Incorrect List Sweep Dwell Time
If the signal generator does not dwell for the correct period of time at each
sweep list point, follow these steps:
1.
Press
Sweep/List
>
Configure List Sweep
.
This displays the sweep list values.
2.