Keysight E6953A V2X Test Solution Installation & User Guide
measured at the RFIN port. Note that any currently configured cable loss
values will be applied.
Figure 3-20
Test Step paths
In the example above, the VXT RFOUT to “V2X DUT” loss was measured as a
4.351232529 dB loss and the “V2X DUT” to VXT RFIN loss was measured as
4.664007664 dB loss. These values should be added into the Cable Loss table.
Figure 3-21
Cable Loss table