2 Programming the Test Set
STATus Subsystem
Initial S/W Revision
Prior to A.02.00
Questionable Event Query
This query returns the decimal value of the sum of the bits in the Questionable Event register.
The register requires that the associated PTR or NTR filters be set before a condition register bit can set a
bit in the event register. The data in this register is latched until it is queried. Once queried, the register is
cleared.
Mode
All
Remote Command
:STATus:QUEStionable[:EVENt]?
Example
STAT:QUES?
Preset
0
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Negative Transition
This command determines which bits in the Questionable Condition register will set the corresponding bit
in the Questionable Event register when the condition register bit has a negative transition (1 to 0). The
variable <integer> is the sum of the decimal values of the bits that you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:NTRansition <integer>
:STATus:QUEStionable:NTRansition?
Example
STAT:QUES:NTR 16
Temperature summary ‘questionable cleared’ will be reported to the Status Byte Register.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Positive Transition
This command determines which bits in the Questionable Condition register will set the corresponding bit
in the Questionable Event register when the condition register bit has a positive transition (0 to 1). The
variable <integer> is the sum of the decimal values of the bits that you want to enable.
Sequence Analyzer User's & Programmer's Reference
71