6 List Sequence Measurements
Source
Step Configuration of Transition Time parameter list (Remote Command Only)
This SCPI command is to configure “Transition Time” parameter array of the whole List Sequencer at one
time. The number of array is same as step number defined in
Number of
Steps. As a step is setup, the value entered run through several levels of validation.
Remote Command
:SOURce:LIST:SETup:TRANsition:TIME <time>, <time>, <time>, …
:SOURce:LIST:SETup:TRANsition:TIME?
Example
:SOUR:LIST:SET:TRAN:TIME 1ms,1ms,1ms
:SOUR:LIST:SET:TRAN:TIME?
Notes
The command is to setup below parameter array of whole list sequence.
Transition Time <
time
> – specifies the transition time for the stepin seconds. For details of the valid
ranges for the transition time see
If input parameter number exceeds the step number defined by
Number of Steps then generate error ", and only those parametes whose index number falls in
number of steps will be updated.
Dependencies
The range is 1 to 1000 which is determined by the number of steps you have configured. For details
see
Initial S/W Revision
A.09.40
Step Configuration of Radio Band parameter list (Remote Command Only)
This SCPI command is to configure “Radio Band” parameter array of the whole List Sequencer at one time.
The number of array is same as step number defined in
Number of Steps.
As a step is setup, the value entered run through several levels of validation.
R
e
m
o
t
e
C
o
m
m
a
n
d
:SOURce:LIST:SETup:RADio:BAND <enum>, <enum>, <enum>, …
:SOURce:LIST:SETup:RADio:BAND?
E
x
a
m
p
l
:SOUR:LIST:SET:RAD:BAND PGSM, EGSM, RGSM
:SOUR:LIST:SET:RAD:BAND?
632
Sequence Analyzer User's & Programmer's Reference