![Keysight E6650A EXF User Programming Manual Download Page 592](http://html1.mh-extra.com/html/keysight/e6650a-exf/e6650a-exf_user-programming-manual_1961647592.webp)
6 List Sequence Measurements
Source
Key Path
Source, Modulation Setup, ARB, Header Utilities
Remote Command
:SOURce:RADio:ARB:HEADer:SAVE
Example
:SOUR:RAD:ARB:HEAD:SAVE
Notes
Attempting to save the header details via SCPI when no waveform was selected for playback will
generate an error.
Initial S/W Revision
A.05.00
Modified at S/W Revision
A.16.00 (AWGN settings will be saved to header as well )
Query Waveform Unique ID (Remote Command Only)
Each Signal Studio waveform contains a unique waveform ID, which recorded in the header. This command
allows you to query the unique waveform ID from the header. This is a SCPI only command. User can also
check
Error! Reference source not found.
for waveform unique ID display.
Remote Command
:MMEMory:HEADer:ID? “<file name>”
Example
:MMEM:HEAD:ID? “test.wfm” (query the waveform already loaded into the ARB memory)
:MMEM:HEAD:ID? “D: VARB\test.wfm” (query the waveform on the hard disk by absolute path)
:MMEM:HEAD:ID? “NVWFM:test.wfm” (query the waveform on the hard disk by MSUS)
Notes
SCPI query only. The queried waveform file can be in ARB memory, or on hard disk. If want to query
ARB in ARB memory, then give out the file name directly. If want to query ARB on the hard disk, then
absolute file path or MSUS should be given along with the file name. The valid MSUS is “NVWFM”
which is mapped to D: VARB on hard disk.
If the file cannot be found in ARB memory or on hard disk, an error is generated.and value –1 is
returned
Initial S/W Revision
A.09.00
Bus Trigger Command (Remote Command Only)
Used to initiate an immediate trigger event if the trigger source is set to Bus.
Remote Command
:SOURce:RADio:ARB:TRIGger:INITiate
Example
:SOUR:RAD:ARB:TRIG:INIT
Initial S/W Revision
A.05.00
AM
Allows access to the menu for configuring the Amplitude Modulation.
592
Sequence Analyzer User's & Programmer's Reference