6 List Sequence Measurements
Meas Setup
LSEQ:INCL:SOUR?
Dependencies
•
If Source->List Sequencer->Sequencer is set OFF, “Include Source” will be disabled.
•
If Source->List Sequencer->Sequencer is set ON, “Include Source” will be enabled.
Couplings
If Source->List Sequencer->Sequencer is set OFF, “Include Source” will be set to NO.
Preset
On
State Saved
Saved in instrument state.
Range
YES|NO|0|1
Initial S/W Revision
A.05.00
Help Map ID
35198
Abort on Limit Fail
This parameter is used to toggle abort on a limit failing while a sequence measurement is being made.
When set to ON, if any of the selected measurements on an analysis step fails its pass/fail limits, the
sequence aborts at the current point and fails to complete the full sequence if any more acquisitions or
measurements have still to be made. When set to OFF, if any measurements fail their limits, then the
sequence acquisitions and measurements carry on regardless. This only applies to measurements that
have their respective limit test parameters enabled.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Abort On Limit Fail
Mode
SEQA
Remote Command
[:SENSe]:LSEQuencer:ABORt:LIMit:FAIL[:STATe] ON|OFF|1|0
[:SENSe]:LSEQuencer:ABORt:LIMit:FAIL[:STATe]?
Example
LSEQ:ABOR:LIM:FAIL
ON
LSEQ:ABOR:LIM:FAIL?
Dependencies
This only applies to limits that have their respective limit test parameters enabled.
Measurements that this parameter does not apply to due to not having limits are:
•
Basic Discrete PAvT
•
Basic IQ Data
•
WCDMA Code Domain Power
Preset
OFF
State Saved
Saved in instrument state.
Range
ON | OFF | 1 | 0
Initial S/W Revision
A.05.00
Help Map ID
35199
Sequence Analyzer User's & Programmer's Reference
465