6 List Sequence Measurements
Meas Setup
[:SENSe]:LSEQuencer:BFERor:DIF:BANDwidth[:RESolution]?
Example
LSEQ:BFER:DIF:BAND 25 MHz
LSEQ:BFER:DIF:BAND?
Couplings
If Digital IF BW is set differently for different measurements in an acquisition, the test set’s analyzer uses the
highest value of those set. Basic Frequency and Phase Error makes the measurement on the captured samples,
without re-sampling.
Preset
25 MHz
State Saved
Saved in instrument state.
Min
1 kHz
Max
40 MHz
Help Map
ID
35190
Limit: Freq Error
This parameter sets the pass/fail limit for the frequency error result. Units are in parts per million (ppm).
parameter_table_
23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup,
Remote Command
[:SENSe]:LSEQuencer:BFERor:LIMit:PPM <real>
[:SENSe]:LSEQuencer:BFERor:LIMit:PPM?
[:SENSe]:LSEQuencer:BFERor:LIMit:PPM OFF|ON|0|1
[:SENSe]:LSEQuencer:BFERor:LIMit:PPM STATe?
Example
LSEQ:BFER:LIM:PPM 0.06
LSEQ:BFER:LIM:PPM?
LSEQ:BFER:LIM:PPM:STAT
LSEQ:BFER:LIM:PPM:STAT?
Couplings
If Abort on Limit Fail = On, and the freq error is more than the value set, the Sequence aborts, without
completing the Sequence.
Preset
0.05
ON
State Saved
Saved in instrument state.
Min
0
Max
100
Initial S/W Revision
A.05.00
Help Map ID
35193
458
Sequence Analyzer User's & Programmer's Reference