6 List Sequence Measurements
Meas Setup
Example
LSEQ:BTXP:LIM:UPP 9dB
LSEQ:BTXP:LIM:UPP?
LSEQ:BTXP:LIM:UPP:STAT
LSEQ:BTXP:LIM:UPP:STAT?
Couplings
This parameter is also coupled with amplitude correction if it is enabled.
Preset
9
ON
State Saved
Saved in instrument state.
Min
–100
Max
100
Initial S/W Revision
A.05.00
Help Map ID
35187
Lower Limit
This parameter sets the threshold level for the Basic Tx Power pass/fail result. When set to ON, the
measurement checks each Analysis Step relative to the “Exp Power” level set for each Analysis Step. If it is
below the “Exp Power” level plus this lower limit value, then the Pass/Fail result is “1”. If above, the
Pass/Fail result is “0”. When set to OFF, the measurement populates the Pass/Fail result with “–1”,
signifying it’s untested.
This parameter is also coupled with amplitude correction if it is enabled.
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Basic Meas Setup, Basic Transmit Power Setup
Remote Command
[:SENSe]:LSEQuencer:BTXPower:LIMit:LOWer <rel_ampl>
[:SENSe]:LSEQuencer:BTXPower:LIMit:LOWer?
[:SENSe]:LSEQuencer:BTXPower:LIMit:LOWer:STATe OFF|ON|0|1
[:SENSe]:LSEQuencer:BTXPower:LIMit:LOWer:STATe?
Example
LSEQ:BTXP:LIM:LOW 9dB
LSEQ:BTXP:LIM:LOW?
LSEQ:BTXP:LIM:LOW:STAT
LSEQ:BTXP:LIM:LOW:STAT?
Couplings
This parameter is also coupled with amplitude correction it is enabled.
Preset
–10
ON
State Saved
Saved in instrument state.
Min
–100
Max
100