6 List Sequence Measurements
Meas Setup
parameter_table_23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup
Notes
No SCPI, front panel access only
Cannot delete when
Number of Acquisitions
is equal to 1.
Couplings
Decreases
Number of Acquisitions
by 1
Initial S/W Revision
A.05.00
Help Map ID
35030
Avg Number
Sets the number of averages for the acquisition.
parameter_
table_
23.52941
52.94118
Key Path
Meas Setup, Acquisition Setup
Remote
Command
- To set average number individually,
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:AVERage:NUMBer <integer>
[:SENSe]:LSEQuencer:ACQuire{1:512}:SETup:AVERage:NUMBer?
-To set the average number as an array:
[:SENSe]:LSEQuencer:LIST:SETup:AVERage:NUMBer <integer>,<integer>,<integer>,….
[:SENSe]:LSEQuencer:LIST:SETup:AVERage:NUMBer?
The array form of the command is used to make a different setting for each acquisition in the sequence (the
number of array values must match the number of acquisitions; otherwise a "settings conflict" error will occur).
Example
:LSEQ:ACQ2:SET:AVER:NUMB 3
:LSEQ:ACQ2:SET:AVER:NUMB?
:LSEQ:LIST:SET:AVER:NUMB 2,3,4,5
:LSEQ:LIST:SET:AVER:NUMB?
Notes
For setting this parameter using SCPI, see
"Programming Acquisitions Via SCPI" on page 376
.
Preset
1
State Saved
Saved in instrument state.
Range
1 to 512
Min
1
Max
512
Initial S/W
A.05.00
Sequence Analyzer User's & Programmer's Reference
385