5 System Functions
System
FEC Self Test Results
Provides a display of last FEC test results, the display should appear listing model number, serial number
and test time at the top of display, and then list test date/time, test name, measured value, valid range
and pass/fail of each FEC test item, the tabular data should be directly printable.
Key Path
System, Diagnostics, Self Test,Show Results
Remote Command
SYSTem:TEST:WCTS:SHOW:RESult FEC
Example
SYST:TEST:WCTS:SHOW:RES FEC
Initial S/W Revision
A.12.50
The example of FEC self test result display is as follows:
FEC Self Test Results
Produce Number: E6607C
Serial Number: MY51380437
Instrument S/W: 11/16/2012 2:51:19 PM
FpgaVersionTest
Date
Time(GMT)
Name
MeasValue
ValidRange
Result
11/23/2012
16:13:56
Analog_FPGA
16.000
>= 16.000
Pass
11/23/2012
16:13:56
Digital_FPGA
50.000
>= 46.000
Pass
11/23/2012
16:13:56
CRFS_FPGA
38.000
>= 38.000
Pass
PowerSupplyTest
Date
Time(GMT)
Name
MeasValue
ValidRange
Result
11/23/2012
16:13:56
ABUS_+32CHK
31.904
30.900 –
32.900
Pass
11/23/2012
16:13:56
ABUS_+12CHK
12.296
10.800 –
13.200
Pass
11/23/2012
16:13:56
+10VA
9.935
9.600 –
10.200
Pass
11/23/2012
16:13:56
+5VA
4.995
4.900 – 5.100 Pass
11/23/2012
16:13:56
+3.3VA
3.299
3.200 – 3.400 Pass
11/23/2012
16:13:56
–3.3VA
–3.311
–3.400 - –
3.200
Pass
11/23/2012
16:13:56
ACOM
0.00
–0.200 –
0.200
Pass
11/23/2012
16:13:56
–5VA
–5.036
–5.100 - –
4.900
Pass
11/23/2012
16:13:56
–6.1VA
–5.880
–6.200 - –
5.700
Pass
204
Sequence Analyzer User's & Programmer's Reference