5 System Functions
System
Key Path
System, Restore System Defaults
Example
:SYST:DEF MISC
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
x.14.50
Restore Mode Defaults (All Modes)
This selection resets all of the modes in the instrument back to their default state just as a Restore Mode
Defaults does and it switches the instrument to the power-on mode and causes the default measurement
for the power-on mode to be active. This level of Restore System Defaults does not affect any system
settings, but it does affect the state of all modes and does cause a mode switch unless the instrument was
already in the power-on mode.
Confirmation is required to restore the factory default values. The confirmation dialog is:
Key Path
System, Restore System Defaults
Example
:SYST:DEF MOD
Couplings
An All Mode will cause the currently running measurement to be aborted, mode switch to the power-
on mode and activate the default measurement for the power-on mode.. It gets the mode to a
consistent state with all of the default couplings set.
Initial S/W Revision
Prior to A.02.00
All
This performs a comprehensive reset of ALL analyzer settings to their factory default values. It resets all of
the system setting groups, causes a Restore Mode Defaults for all modes in the instrument, and switches
back to the power-on mode. It does not affect the User Preset file or any user saved files.
Confirmation is required to restore the factory default values. The confirmation dialog is:
194
Sequence Analyzer User's & Programmer's Reference