3 Input/Output Functions
Input/Output
Key Path
Input/Output, Corrections
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|6:DELete
Example
CORR:CSET:DEL
CORR:CSET1:DEL
CORR:CSET4:DEL
Notes
Pressing this key when no corrections are present is accepted without error.
Initial S/W Revision
A.02.00
Apply Corrections
Applies amplitude corrections, which are marked as ON to the measured data. If this is set to OFF, then no
amplitude correction sets will be used, regardless of their individual on/off settings. If set to ON, the
corrections that are marked as ON (see
"Correction On/Off" on page 102
) are used.
Key Path
Input/Output, Corrections
Remote Command
[:SENSe]:CORRection:CSET:ALL[:STATe] ON|OFF|1|0
[:SENSe]:CORRection:CSET:ALL[:STATe]?
Example
SENS:CORR:CSET:ALL OFF
This command makes sure that no amplitude corrections are applied, regardless of their individual
on/off settings.
Preset
Not affected by Preset. Set to OFF by Restore Input/Output Defaults
State Saved
Saved in instrument state.
Initial S/W Revision
A.02.00
Delete All Corrections
Erases all correction values for all 4 Amplitude Correction sets.
When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to
delete all corrections. Press ESC or Cancel to close this dialog.” The deletion is only performed if you
press OK or Enter.
Key Path
Input/Output, Corrections
Remote Command
[:SENSe]:CORRection:CSET:ALL:DELete
Example
CORR:CSET:ALL:DEL
Initial S/W Revision
A.02.00
112
Sequence Analyzer User's & Programmer's Reference