![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 2673](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616452673.webp)
15 Power Control Measurement
Meas
State Saved
Saved in instrument state.
Range
OFF|ON|0|1
Initial S/W Revision
Prior to A.02.00
RRC Filter State
Toggles the root-raised cosine filter function between On and Off.
Mode
WCDMA
Remote Command
[:SENSe]:PCONtrol:FILTer[:RRC][:STATe] 0|1|OFF|ON
[:SENSe]:PCONtrol:FILTer[:RRC][:STATe]?
Example
PCON:FILT OFF
PCON:FILT?
Notes
You must be in the W-CDMA mode to use this command. Use INSTrument:SELect to set the mode.
Dependencies
When Meas Method is Waveform, this key is grayed out.
Couplings
When RRC Filter Auto is ON, RRC Filter State changes depending on the Meas Method: When Meas
Method is changed to Chip Power, RRC Filter State is set to ON; and when Meas Method is changed
to Waveform, it is set to OFF.
Preset
0
State Saved
Saved in instrument state.
Range
On|Off
Initial S/W Revision
Prior to A.02.00
Capture Interval
Accesses a menu that enables you to select the number of frames for data acquisition.
Key Path
Meas Setup
Mode
WCDMA
Notes
For consistency, this parameter must be controlled with a float numeric from remote commands, but
the front panel interface uses Enum for easier operation.
Dependencies
This Parameter is coupled with MeasType. 8 Frames cannot be set (menu key is grayed out) when
MeasType is SlotPhase.
Couplings
See Dependencies
Preset
4 Frames
State Saved
Saved in instrument state.
Range
1 Frame|2 Frames|4 Frames|8 Frames
Initial S/W Revision
Prior to A.02.00
W-CDMA & HSPA Measurement Applications Reference
2673
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...