15 Power Control Measurement
Meas
Preset
2368.0
State Saved
Saved in instrument state.
Min
1.0
Max
:CALCulate:PCONtrol:SLOT:LENGth –:CALCulate:PCONtrol:SLOT:OFFSet
Initial S/W Revision
Prior to A.02.00
PCG Length
Specifies the number of chips to be used as the integration time for the slot power measurement.
Key Path
Meas Setup, Slot Power Meas
Mode
WCDMA
Remote Command
:CALCulate:PCONtrol:SLOT:LENGth <real>
:CALCulate:PCONtrol:SLOT:LENGth?
Example
CALC:PCON:SLOT:LENG 2560.0
CALC:PCON:SLOT:LENG?
Notes
You must be in the W-CDMA mode to use this command. Use INSTrument:SELect to set the mode.
Preset
2560.0
State Saved
Saved in instrument state.
Min
:CALCulate:PCONtrol:SLOT:DELay + :CALCulate:PCONtrol:SLOT:INTerval or 768.0 (larger one)
Max
25600.0
Initial S/W Revision
Prior to A.02.00
Meas Method
Enables you to access the Meas Method Menu to select the data acquisition method, Waveform or Chip
Power.
•
WAVeform–Measures and performs a power calculation based on the waveform (raw data of A/D). This
means a raw measurement under the specific BW, filter type and Capture Interval.
•
CPOWer–Measures and performs a power calculation based on the chip power (resample on chip
clock). This means a re-sampled power measurement based on the chip clock timing for W-CDMA.
The Meas Method is available only for the Slot Power and PRACH Power measurement types.
In the Slot Phase measurement type, resampling on the chip clock is always performed.
Key Path
Meas Setup
Mode
WCDMA
Remote Command
[:SENSe]:PCONtrol:METHod WAVeform|CPOWer
[:SENSe]:PCONtrol:METHod?
W-CDMA & HSPA Measurement Applications Reference
2671
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...