![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 2035](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616452035.webp)
13 Modulation Accuracy (Composite EVM) Measurement
Marker
Couple Markers
Toggles the state of the markers to be coupled On or Off. When this function is true (On), moving any
marker causes an “equal X Axis movement” of every other marker that is active. By “equal X Axis
movement” we mean that we preserve the difference between each marker’s X Axis value (in the
fundamental x-axis units of the trace that marker is on) and the X Axis value of the marker being moved (in
the same fundamental x-axis units).
This may result in markers going offscreen.
Key Path
Marker
Mode
WCDMA
Remote Command
:CALCulate:RHO:MARKer:COUPle[:STATe] ON|OFF|1|0
:CALCulate:RHO:MARKer:COUPle[:STATe]?
Example
CALC:RHO:MARK:COUP ON
CALC:RHO:MARK:COUP?
Notes
When the marker is assigned to the IQ Measured Polar graph, a Chip value is coupled instead of an X
Axis value.
Preset
OFF
State Saved
Saved in instrument state.
Range
On|Off
Initial S/W Revision
Prior to A.02.00
All Markers Off
Turns off all markers.
Key Path
Marker
Mode
WCDMA
Remote Command
:CALCulate:RHO:MARKer:AOFF
Example
CALC:RHO:MARK:AOFF
Initial S/W Revision
Prior to A.02.00
Marker Chip Value (Remote Command only)
Sets the marker Chip value in the current marker for the I/Q Polar trace. It has no effect if the control mode
is Off, but if the control mode is Normal, this is the SCPI equivalent of entering a Chip value.
This command is valid only when Marker Trace ‘POLar’(I/Q Polar)is active. For any other Marker Trace, the
command is ignored.
Mode
WCDMA
Remote Command
:CALCulate:RHO:MARKer[1]|2|...|12:CHIP <real>
W-CDMA & HSPA Measurement Applications Reference
2035
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...