4 Input/Output Functions
Input/Output
which DOES contain an antenna unit, a Mass Storage error is generated.
This command will generate an “Option not available” error unless you have the proper option
installed in your instrument.
Preset
Not affected by a Preset. Set to OFF by Restore Input/Output Defaults
State Saved
Saved in instrument state.
Backwards Compatibility
Notes
Unlike legacy analyzers, Preset does not turn Corrections off (Restore Input/Output Defaults does).
Initial S/W Revision
A.02.00
Modified at S/W Revision
A.14.00
Properties
Accesses a menu that lets you set the properties of the selected correction.
Key Path
Input/Output, Corrections
Initial S/W Revision
A.02.00
Select Correction
Specifies the selected correction. The term "selected correction" is used throughout this document to
specify which correction will be affected by the functions.
Key Path
Input/Output, Corrections
Mode
SA, I/Q Analyzer, Phase Noise, VXA, RTSA, EMI Receiver, DVB-T/H, DTMB, DVB-T/H, DTMB, W-
CDMA, LTE & LTE-Adv FDD, LTE & LTE-Adv TDD, Sequence Analyzer, BTooth
Notes
The selected correction is remembered even when not in the correction menu.
Preset
Set to Correction 1 by Restore Input/Output Defaults
Readback
Correction 1|Correction 2|Correction 3|Correction 4|Correction 5|Correction
6|Correction7|Correction8
Initial S/W Revision
A.02.00
Modified at S/W Revision
x.14.50
Antenna Unit
For devices (like antennas) that make measurements of field strength or flux density, the correction array
should contain within its values the appropriate conversion factors such that, when the data on the
analyzer is presented in dBµV, the display is calibrated in the appropriate units. The "Antenna Unit" used
for the conversion is contained within the corrections array database. It may be specifiedor loaded in from
an external file or SCPI.
W-CDMA & HSPA Measurement Applications Reference
203
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...