4 Input/Output Functions
Input/Output
Key Path
Input/Output, RF Input
Remote Command
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude] 50|75
[:SENSe]:CORRection:IMPedance[:INPut][:MAGNitude]?
Example
CORR:IMP 75 sets the input impedance correction to 75 ohms.
CORR:IMP?
Preset
This is unaffected by a Preset but is set to 50 ohms on a "Restore Input/Output Defaults" or "Restore
System Defaults->All"
Some instruments/options may have 75 ohms available.
State Saved
Saved in instrument state
Readback
50
Ω
or 75
Ω
. Current setting reads back to the RF key.
Initial S/W Revision
Prior to A.02.00
RF Input Port
Specifies the RF input port used. The RF Input Port key only appears on units with multiple inputs, and lets
you switch between the two inputs.
Switching from the RF input port to one of the RFIO ports, on units that have them, changes the receiver
performance of the instrument.
Key Path
Input/Output, RF Input
Remote Command
[:SENSe]:FEED:RF:PORT[:INPut] RFIN | RFIN2 | RFIO1 | RFIO2 | RFIO3 |
RFIO4
[:SENSe]:FEED:RF:PORT[:INPut]?
Example
:FEED:RF:PORT RFIN
Dependencies
This key only appears in models that support multiple inputs. If the SCPI command is sent with
unsupported parameters in any other model, an error is generated, –221.1900, “Settings
conflict;option not installed”
When any input is selected in a measurement that does not support it, the "No result; Meas invalid
with this input" error condition occurs, and the measurement returns invalid data when queried.
Preset
This is unaffected by Mode Preset but is set to RF on a "Restore Input/Output Defaults" or "Restore
System Defaults -> All"
State Saved
Saved in instrument state
Readback
The current RF Input Port selected is read back to this key
Backwards
Compatibility SCPI
INPut<1|2>:TYPE INPUT1 | INPUT2
INPut<1|2>:TYPE?
Included for R&S ESU compatibility. In the MXE, the INPUT1 parameter is aliased to RFIN and the
INPUT2 parameter is aliased to RFIN2
Initial S/W Revision
A.05.01
Modified at S/W Revision
A.14.00
W-CDMA & HSPA Measurement Applications Reference
193
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...