![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1754](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451754.webp)
12 Code Domain Measurement
Meas Setup
[:SENSe]:CDPower:FILTer:ALPHa?
Example
CDP:FILT:ALPH 0.3
CDP:FILT:ALPH?
Preset
0.22
State Saved
Saved in instrument state.
Min
0.01
Max
0.50
Backwards
Compatibility SCPI
[:SENSe]:CDPower:ALPHa
Initial S/W Revision
Prior to A.02.00
Chip Rate
Changes the chip rate.
Key Path
Meas Setup, Advanced
Mode
WCDMA
Remote Command
[:SENSe]:CDPower:CRATe <freq>
[:SENSe]:CDPower:CRATe?
Example
CDP:CRAT 4000000
CDP:CRAT?
Preset
3.84 MHz
State Saved
Saved in instrument state.
Min
3.456 MHz
Max
4.224 MHz
Initial S/W Revision
Prior to A.02.00
Multi Channel Estimator
Toggles the multi channel estimator function by MMSE between On and Off.
•
On: The individual code channels are aligned to the pilot channel to improve the phase error (whether
each code phase is aligned or not). This takes a longer time.
•
Off: The phase information is computed from one coded signal only. (The phase of each code channel
needs to be aligned to the pilot channel.)
Key Path
Meas Setup, Advanced
Mode
WCDMA
Remote Command
[:SENSe]:CDPower:MCEStimator OFF|ON|0|1
[:SENSe]:CDPower:MCEStimator?
1754
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...