![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 159](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_1961645159.webp)
3 Programming the Test Set
STATus Subsystem
Questionable Calibration Skipped Negative Transition
This command determines which bits in the Questionable Calibration Skipped Condition register will set
the corresponding bit in the Questionable Calibration Skipped Event register when the condition register bit
has a negative transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:SKIPped:NTRansition <integer>
:STATus:QUEStionable:CALibration:SKIPped:NTRansition?
Example
STAT:QUES:CAL:SKIP:NTR 1 Align RF skipped is not required.
Preset
0
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Skipped Positive Transition
This command determines which bits in the Questionable Calibration Skipped Condition register will set
the corresponding bit in the Questionable Calibration Skipped Event register when the condition register bit
has a positive transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that you
want to enable.
Mode
All
Remote Command
:STATus:QUEStionable:CALibration:SKIPped:PTRansition <integer>
:STATus:QUEStionable:CALibration:SKIPped:PTRansition?
Example
STAT:QUES:CAL:SKIP:PTR 1 Align RF skipped is required.
Preset
32767
Min
0
Max
32767
Status Bits/OPC
dependencies
Sequential command
Initial S/W Revision
Prior to A.02.00
Questionable Calibration Extended Failure Register
"Questionable Calibration Extended Failure Condition " on page 160
"Questionable Calibration Extended Failure Enable " on page 160
"Questionable Calibration Extended Failure Event Query " on page 160
"Questionable Calibration Extended Failure Negative Transition " on page 161
W-CDMA & HSPA Measurement Applications Reference
159
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...