![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1370](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451370.webp)
11 Occupied Bandwidth Measurement
Meas
Remote Command
:CALCulate:DATA[n]?
Notes
The return trace depends on the measurement.
In CALCulate:<meas>:DATA[n], n is any valid subopcode for the current measurement. It returns the
same data as the FETCh:<measurement>? query where <measurement> is the current
measurement.
Initial S/W Revision
Prior to A.02.00
Calculate/Compress Trace Data Query (Remote Command Only)
Returns compressed data for the currently selected measurement and sub-opcode [n].
n = any valid sub-opcode for that measurement. See the MEASure:<measurement>? command description
of your specific measurement for information on the data that can be returned.
The data is returned in the current Y Axis Unit of the analyzer. The command is used with a sub-opcode <n>
(default=1) to specify the trace. With trace queries, it is best if the analyzer is not sweeping during the
query. Therefore, it is generally advisable to be in Single Sweep, or Update=Off.
This command is used to compress or decimate a long trace to extract and return only the desired data. A
typical example would be to acquire N frames of GSM data and return the mean power of the first burst in
each frame. The command can also be used to identify the best curve fit for the data.
Remote Command
:CALCulate:DATA<n>:COMPress? BLOCk | CFIT | MAXimum | MINimum | MEAN |
DMEan | RMS | RMSCubed | SAMPle | SDEViation | PPHase [,<soffset>
[,<length>[,<roffset>[,<rlimit>]]]]
Example
To query the mean power of a set of GSM bursts:
Supply a signal that is a set of GSM bursts.
Select the IQ Waveform measurement (in IQ Analyzer Mode).
Set the sweep time to acquire at least one burst.
Set the triggers such that acquisition happens at a known position relative to a burst.
Then query the mean burst levels using, CALC:DATA2:COMP? MEAN, 24e–6, 526e–6 (These
parameter values correspond to GSM signals, where 526e–6 is the length of the burst in the slot and
you just want 1 burst.)
Notes
The command supports 5 parameters. Note that the last 4 (<soffset>,<length>,<roffset>,<rlimit>)
are optional. But these optional parameters must be entered in the specified order. For example, if
you want to specify <length>, then you must also specify <soffset>. See details below for a definition
of each of these parameters.
This command uses the data in the format specified by FORMat:DATA, returning either binary or
ASCII data.
Initial S/W Revision
Prior to A.02.00
•
BLOCk or block data - returns all the data points from the region of the trace data that you specify. For
example, it could be used to return the data points of an input signal over several timeslots, excluding
the portions of the trace data that you do not want. (This is x,y pairs for trace data and I,Q pairs for
complex data.)
1370
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...