![Keysight E6640A Measurement Applications User & Programmer Reference Download Page 1024](http://html1.mh-extra.com/html/keysight/e6640a/e6640a_measurement-applications-user-and-programmer-reference_19616451024.webp)
10 Spectrum Emission Mask Measurement
Marker
:CALCulate:SEMask:MARKer[1]|2|...|12:X?
Example
CALC:SEM:MARK3:X 1.0 GHz
CALC:SEM:MARK3:X?
Notes
If no suffix is sent it will use the fundamental units for the current marker X Axis Scale. If a suffix is
sent that does not match the current marker X Axis Scale unit, an error “Invalid suffix” will be
generated.
The query returns the marker’s absolute X Axis value if the control mode is Normal. The query is
returned in the fundamental units for the current marker X Axis scale. If the marker is Off the
response is not a number.
When a Marker is turned on, it is placed center of the screen on the trace. Therefore the default value
depends on instrument condition, although the Preset/Default is defined as 1.5 GHz.
Preset
After a preset, all Markers are turned OFF, so a Marker X Axis Value query will return a not a number
(NAN).
State Saved
No
Min
–9.9E+37
Max
9.9E+37
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.02.00, A.03.00
Marker X Axis Position (Remote Command Only)
Sets the marker X position in trace points. It has no effect if the control mode is Off, but is the SCPI
equivalent of entering a value if the control mode is Normal, except in trace points rather than X Axis Scale
units. The entered value is immediately translated into the current X Axis Scale units for setting the value
of the marker.
Mode
SA, WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, DTMB (CTTB), DVB-T/H, ISDB-T, CMMB,
LTE, LTETDD, Digital Cable TV,
WLAN, MSR, LTEAFDD, LTEATDD
Remote Command
:CALCulate:SEMask:MARKer[1]|2|...|12:X:POSition <real>
:CALCulate:SEMask:MARKer[1]|2|...|12:X:POSition?
Example
CALC:SEM:MARK10:X:POS 1001
CALC:SEM:MARK10:X:POS?
Notes
The query returns the marker’s absolute X Axis value in trace points if the control mode is Normal.
The value is returned as a real number, not an integer, corresponding to the translation from X Axis
Scale units to trace points . If the marker is Off the response is not a number.
When a Marker is turned on, it is placed center of the screen on the trace. Therefore the default value
depends on the instrument condition although the Preset/Default is defined as 6507 (this value
might be the expected value when all the offsets are on).
Preset
After a preset, all Markers are turned OFF, so a Marker X Axis Value query will return a not a number
(NAN).
State Saved
No
1024
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...