10 Spectrum Emission Mask Measurement
AMPTD Y Scale
Range
Represents the amplitude of the largest sinusoidal signal that could be present within the IF without being
clipped by the ADC. For signals with high peak-to-average ratios, the range may need to exceed the rms
signal power by a fair amount to avoid clipping.
Key Path
Range
Mode
BASIC
Remote Command
[:SENSe]:POWer[:RF]:RANGe <real>
[:SENSe]:POWer[:RF]:RANGe?
Example
:POW:RANG 10.0
:POW:RANG?
Notes
The MIN and MAX values are affected by the External Gain parameters, and by the Center Frequency.
(The hardware compensates for frequency response and alters the Range setting.)
Preset
0
State Saved
Saved in instrument state.
Min
–100
Max
100
Initial S/W Revision
A.12.50
Adjust Range For Min Clip
Sets the combination of attenuation and gain based on the current measured signal level so that clipping
will be at a minimum.
This is an "immediate action" function, that is, it executes once, when the key is pressed.
This key is grayed out in measurements that do not support this functionality.
Key Path
AMPTD Y Scale, Attenuation
Remote Command
[:SENSe]:POWer[:RF]:RANGe:OPTimize IMMediate
Initial S/W Revision
Prior to A.02.00
Modified at S/W Revision
A.03.00
Pre-Adjust for Min Clip
If this function is on, it does the adjustment described under Adjust Range For Min Clip each time a
measurement restarts. Therefore, in Continuous measurement mode, it only executes before the first
measurement.
Key Path
AMPTD Y Scale, Attenuation
Remote Command
[:SENSe]:POWer[:RF]:RANGe:OPTimize:ATTenuation OFF | ON | ELECtrical |
COMBined
1004
W-CDMA & HSPA Measurement Applications Reference
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...