Keysight E5063A Network Analyzer 27
Troubleshooting
Troubleshooting Using Diagnostic Test
Troubleshooting Using Diagnostic Test
The E5063A has diagnostic test function to diagnose the analog measurement
section and internal dc power supply voltages. The diagnostic tests make it
possible to isolate a faulty board assembly. The following paragraphs describe
the procedure to perform the diagnostic tests.
Diagnostic Test
To execute the diagnostic test
To isolate faulty board assembly modules, execute the diagnostic test in
accordance with the following procedure.
Step 1.
Press
key.
Step 2.
Click
Service Menu
>
Test Menu
>
Diagnostic Test
.
Figure 1-12
Diagnostic Test 1/3
To perform the diagnostic test properly, the following conditions must be
met:
Environmental temperature: 23°C ±5°C
Warm up time: > 30 minutes
Do not operate front panel keys, keyboard and mouse during the
diagnostic test. Changing the instrument setting while the diagnostic test
is in progress will cause incorrect test results.