Remote Interface Reference
3
Halting an Output in Progress
E3632A User’s Guide
121
Halting an Output in Progress
You can send a device clear at any time to stop an output in
progress over the GPIB interface. The status registers, the
error queue, and all configuration states are left unchanged
when a device clear message is received. Device clear
performs the following actions.
•
The power supply’s input and output buffers are cleared.
•
The power supply is prepared to accept a new command
string.
•
The following statement shows how to send a device clear
over the GPIB interface using Keysight BASIC.
CLEAR 705
IEEE- 488 device clear
•
The following statement shows how to send a device clear
over the GPIB interface using the GPIB Command Library
for C or QuickBASIC.
IOCLEAR (705)
N O T E
For RS-232 operation, sending the
<Ctrl-C>
character will perform the
same operation as the IEEE-488 device clear message. The power supply’s
DTR (data terminal ready) handshake line is set true following a device
clear message. Refer to
“DTR/DSR handshake protocol” on page 60
for
more information.
N O T E
All remote interface configurations can be entered only from the front
panel. Refer to
“Storing and Recalling Operating States” on page 33
to
configure for GPIB or RS-232 interface before operating the power supply
remotely.
Summary of Contents for E3632A
Page 1: ...Keysight E3632A DC Power Supply User s Guide ...
Page 2: ......
Page 3: ......
Page 12: ...X E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...XVIII E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...XX E3632A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...