Calibration Procedures
3
Keysight E3631A Service Guide
85
Calibration record for keysight E3631A
Step
Calibration Description
Measurement
Mode (DVM)
Supply being
Adjusted
1
Unsecure the power supply (see page 65).
2
Turn on "CAL MODE"(simultaneously press "Calibrate" and
"Power" keys) until a beep is heard.
3
Move down menu to "CAL SETUP 1" (press "Calibrate" key).
+6 supply
4
"V LO 0.1000 V" appears on the display (change the display to
match the DVM reading; then press ``Calibrate'' key).
V
+6 supply
5
"V HI +5.7000 V" appears on the display (change the display to
match the DVM reading; then press "Calibrate" key).
V
+6 supply
6
"CAL SETUP 2" now appears on the display (press "Calibrate"
key).
+6 supply
7
"I LO 0.100 A" appears on the display (change the display to
match the computed current through 0.1? resistor; then press
"Calibrate" key).
A
+6 supply
8
"I HI +4.750 A" appears on the display (change the display to
match the computed current through 0.1? resistor; then press
"Calibrate" key).
A
+6 supply
9
"CAL SETUP 3" now appears on the display (press "Calibrate"
key).
+25 supply
10
"V LO 0.5000 V" appears on the display (change the display to
match the DVM reading; then press "Calibrate" key).
V
+25 supply
11
"V HI +24.750 V" appears on the display (change the display to
match the DVM reading; then press ``Calibrate'' key).
V
+25 supply
12
"CAL SETUP 4" now appears on the display (press "Calibrate"
key).
+25 supply
13
"I LO 0.100 A" appears on the display (change the display to
match the computed current through 0.1? resistor; then press
"Calibrate" key).
A
+25 supply
14
"I HI +0.950 A" appears on the display (change the display to
match the computed current through 0.1? resistor; then press
"Calibrate" key).
A
+25 supply
Summary of Contents for E3631A
Page 1: ...Keysight E3631A Triple Output DC Power Supply Service Guide ...
Page 18: ...18 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 22: ...Keysight E3631A Service Guide 22 THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 24: ...24 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 26: ...26 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 33: ...Keysight E3631A Triple Output DC Power Supply Service Guide 31 2 Quick Start ...
Page 50: ...2 Quick Start 48 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 67: ...Calibration Procedures 3 Keysight E3631A Service Guide 65 Figure 3 5 Transient response time ...
Page 101: ...Theory of Operation 4 Keysight E3631A Service Guide 97 Block Diagram ...
Page 122: ...5 Service 118 Keysight E3631A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...