Turn-On Checkout Procedure
1
Connect line cord to power source and turn LINE switch
on.
2
Push +6V METER switch
and, with no load connected, vary +6 V VOLTAGE
control
over its range and check that the voltmeter responds to the control
setting and the ammeter indicates zero.
3
Set the +6V VOLTAGE control to 6 volts and short the +6 V output terminal
to COM (common) terminal
with an insulated test lead. The ammeter
should indicate a short-circuit output current of approximately 1.0 A. Remove
the short from the output terminals.
4
Push the +20 V METER switch
and turn Tracking Ratio control
fully
clockwise to the Fixed position. With no load connected, vary ±20 V VOLTAGE
control
over its range and check that the voltmeter responds to the control
setting and the ammeter indicates zero.
5
Set the ±20 V VOLTAGE control
to 20 volts and short the +20 V output
terminal
to the COM terminal
with an insulated test lead. The
ammeter should indicate a short-circuit output current of 0.55 A ± 5%.
Remove the short from the output terminals.
6
Repeat steps (4) and (5) for –20 V output.
7
Adjust the +20 V output to 20 volts. Then push –20 V METER switch and check
the effect of the Tracking Ratio control on the voltage of the –20 V output. The
–20 V output should be adjustable from less than 0.5 volts to a maximum of
19 to 21 volts.
CAUTION
Before applying power to the supply, check the label on the heat sink to
make certain that the supply's line voltage option agrees with the line
voltage to be used. If the option does not correspond to your line voltage,
Line voltage option conversion
in the service section
before applying power.
1
2
5
8
9
3
7
6
6
10
9
Summary of Contents for E3630A
Page 1: ...Triple Output Power Supply Keysight Model E3630A Operating and Service Manual ...
Page 10: ...10 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 64: ...6 Component Locator Diagrams 64 Keysight E3630A Operating and Service Manual Main Board Top ...
Page 68: ...6 Component Locator Diagrams 68 Keysight E3630A Operating and Service Manual 7 Segment Top ...
Page 69: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 69 7 Segment Bottom ...