Keysight E1412A Digital Multimeter User Guide
clxxvii
[SENSe:]
Multimeter Command Reference
–
The resolution with the smallest value (i.e., the best resolution) for the
selected range if
MIN
is specified.
–
The resolution with the largest value (i.e., the worst resolution) for the
selected range if
MAX
is specified.
Example
Query the Resolution
VOLT:AC:RANG 100E-03
Set range to 0.1 volts.
VOLT:AC:RES 1.0E-07
Set resolution to 0.1
V.
VOLT:AC:RES?
Query multimeter to return the present resolution.
enter statement
Enter value into computer.
VOLTage[:DC]:APERture
[SENSe:]VOLTage[:DC]:APERture <
number
>
sets the integration time in
seconds for dc voltage measurements. Values are rounded up to the nearest
aperture time shown in the following table.
Parameters
Comments
–
MIN
sets the aperture time to 0.333ms.
MAX
sets the aperture time
to 1.66667 seconds (60Hz) or 2 seconds (50Hz).
–
Setting the aperture time also sets the integration time in power line cycles
(PLCs) and the resolution. For example, an aperture time of 16.7ms (60Hz
line frequency) sets an integration time of 1 PLC. The corresponding
resolution depends on the function and range you select.
–
The
VOLT:APER
command overrides the results of previously executed
VOLT:NPLC
and
VOLT:RES
commands. The last command executed has
priority.
–
The greater the aperture time, the greater the normal mode rejection (and
the lower the reading rate).
–
Related Commands:
CALibration:LFRrequency
–
*RST
Condition:
VOLT:APER 0.166667
seconds (60Hz) or
VOLT:APER 0.20000
(50Hz)
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
0.333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
seconds