Keysight E1412A 6 1/2 Digit Multimeter Service Guide
63
CALibration Command Summary
Adjustments
Example
Enter a new calibration security code:
CAL:SEC:STAT OFF, HP_E1412
Unsecure with the old code.
CAL:SEC:CODE
the_new_code
Enter a new calibration code
(a maximum of 12 characters).
:SECure:STATe
CALibration:SECure:STATe OFF | ON, <code>
unsecures or secures the multimeter for
calibration. The calibration code must be the code set by the CAL:SEC:CODE
command. The state is stored in non-volatile memory.
Parameters
Comments
–
You can substitute decimal values for the OFF (“ 0” ) and ON (“ 1” )
parameters.
–
The multimeter calibration is secured when shipped from the factory.
The security code is set to "HP_E1412".
*RST
does not change the state.
Example
Set the calibration state to unsecured:
CAL:SEC:STAT OFF, HP_E1412
Unsecure multimeter calibration.
:SECure:STATe?
CALibration:SECure:STATe?
returns a "1" or "0" to show whether the calibration
security state is enabled (1) or disabled (0). The number is sent to the output
buffer.
Example
Query the calibration security state:
CAL:SEC:STAT?
Query multimeter calibration security state
enter statement
Enter value into computer.
Parameter Name
Parameter Type
Range of Values
Default Units
OFF|ON
boolean
OFF |
0
| ON | 1
none
code
discrete
up to 12 characters set by
:SEC:CODE
none
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......