Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Adjustments
Introduction
or 2 year interval may be adequate for less demanding applications. Keysight
does not recommend extending calibration intervals beyond 2 years in any
application.
Whatever calibration interval you select, Keysight recommends that complete
re-adjustment should always be performed at the calibration interval. This will
increase your confidence that the Keysight E1412A will remain in specification for
the next calibration interval. This criteria for readjustment provides the best
measure of the multimeter’s long-term stability. Performance data measured this
way can easily be used to extend future calibration intervals.
Automating Adjustment Procedures
The Keysight E1412A Multimeter can be automatically adjusted under computer
control. With computer control, you can perform the complete adjustment
procedure and performance verification tests in less than 30 minutes. You will
require a programmable multifuction calibrator to do this. The multimeter is
configured to the required function and range using SCPI commands. The
calibration value is programmed into the calibrator and sent to the multimeter.
The adjustment procedure is initiated over the remote interface. The multimeter
must be unsecured prior to initiating the adjustment procedure.
CALibration Command Summary
See the end of this chapter for a reprint of the CALibration command from the
SCPI Command Reference of Chapter 3, Keysight E1412A User’s Manual and
SCPI Programming Guide.
Calibration Security Code
This feature allows you to enter a security code (electronic key) to prevent
accidental or unauthorized calibrations of the multimeter. Your multimeter is
secured when you first receive it. You MUST unsecure the multimeter by entering
the correct security code before you can adjust calibration constants. You
unsecure the multimeter with the following command:
CAL:SEC:STAT OFF, <security code>
–
The security code is set to "HP_E1412" for the C-size multimeter when it is
shipped from the factory.
–
The security code is stored in non-volatile memory and does not change
when power is turned off or following a remote interface reset (*RST).
–
The security code can contain up to 12 characters.
–
IF YOU FORGET your security code, you can disable the security feature
momentarily by the procedure that follows titled "Unsecuring the
Multimeter Without the Security Code".
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......