Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Verification Tests
Performance Verification
AC Current Gain Verification
Connect the AC Standard across the multimter’s I and LO input terminals as
shown in Figure 4-6.
1
Set up the multimeter as follows:
Reset multimeter
*RST
Configure for AC current (1 A)
CONF:CURR:AC 1
Select Slow AC filter
DET:BAND 3
2
Set the AC Standard controls as follows: Output 1 A Frequency 1kHz
3
Perform the measurement using the READ? command.
4
Verify the result is within specifed limits and record the result.
5
Change ranges using CURR:AC:RANG <range> and make a measurement for
each ACI input and range shown in Table 4-4 verifying the result is within
specifed limits. Record the result.
6
Perform the measurement using the READ? command.
7
Verify that the result is within specified limits and record the result.
Table 4-5
AC Current Gain Verification Tests
INPUT
INPUT
FREQUENCY
Keysight
E1412A Range
Error from nominal
24 hour
90 day
1 year
1 A
1 kHz
1 A
± 1.9 mA
± 1.9 mA
± 1.9 mA
2 A
1 kHz
3 A
± 16.8 mA
± 16.8 mA
± 16.8 mA
Figure 4-6
AC Current Gain Test Setup.