Keysight E1412A 6 1/2 Digit Multimeter Service Guide
Verification Tests
Performance Verification
2-Wire Resistance Gain Verification
Disconnect the input to the multimeter HI and LO Sense terminals and
disconnect the calibrator from the multimeter input cable.
1
Connect a short across the HI and LO terminals of the multimeter test leads
(at the calibrator end) as shown in Figure 4-3.
2
Measure the 2-Wire OHMS function test cable resistance as follows:
Configure for 2-Wire Resistance, 100
CONF:RES 100
Set CALCulate function to NULL
CALC:FUNC NULL
Enable the calculate state
CALC:STAT ON
Measure the cable resistance
READ?
Enter the measured value (this first measurement stores the cable
resistance as the NULL OFFSet)
3
Remove the short from the cable and connect the DC Standard across the
Keysight E1412A HI and LO Input terminals.
4
Set the DC Standard output to output a 2-wire 100
(this requires EXTernal
SENSE OFF for the Fluke 5700).
5
Perform the measurement using the READ? command.
6
Verify the result is within specified limits and record the result.
7
Change ranges using RES:RANG <range> and make a measurement for each
2-Wire resistance input and range shown in Table 4-2 verifying the result is
within specified limits. Record the result.
Figure 4-3
Resistance (2-Wire) Gain Verification Test Setup.
Summary of Contents for E1412A
Page 1: ...Keysight E1412A 6 1 2 Digit Multimeter Service Manual 75000 Series C ...
Page 2: ......
Page 6: ...vi ...
Page 12: ...xii Keysight E1412A 6 1 2 Digit Multimeter Service Guide ...
Page 74: ...68 Keysight E1412A 6 1 2 Digit Multimeter Service Guide Adjustments CALibration ...
Page 80: ...74 E1412A 6 1 2 Digit Multimeter Service Guide Backdating ...
Page 118: ......