I2C/SPI Triggering and Serial Decode
28
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
449
Clock Timeout
— Select
Clock Timeout
in the
Frame by
softkey, then select
the
Timeout
softkey and turn the Entry knob to set the minimum time that
the Clock signal must be idle (not transitioning) before the oscilloscope
will search for the Data pattern on which to trigger.
The Timeout value can be set anywhere from 100 ns to 10 s.
When you press the
Frame by
softkey, the
Display Info
graphic changes to
show timeout selection or the current state of the chip select signal.
b
Press the
Threshold
softkey; then, turn the Entry knob to select the chip select
signal threshold voltage level.
The threshold voltage level is used in decoding, and it will become the
trigger level when the trigger type is set to the selected serial decode slot.
When
Display Info
is enabled, information about the selected signal sources and
their threshold voltage levels, as well as a waveform diagram, appears on the
screen.
SPI Triggering
To set up the oscilloscope to capture SPI signals, see
After the oscilloscope has been set up to capture SPI signals, you can then trigger
on a data pattern that occurs at the start of a frame. The serial data string can be
specified to be from 4 to 32 bits long.
When you select the SPI trigger type and
Display Info
is enabled, a graphic is
displayed showing the current state of the frame signal, clock slope, number of
data bits, and data bit values.
1
Press
[Trigger]
.
2
In the Trigger Menu, press the
Trigger Type
softkey; then, turn the Entry knob to
select the serial slot (Serial 1 or Serial 2) on which the SPI signals are being
decoded.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...