428
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
27
FlexRay Triggering and Serial Decode
5
Press
Source
and select the analog channel that is probing the FlexRay signal.
6
Press
Threshold
; then, turn the Entry knob to set the threshold voltage level.
The threshold level should be set below the idle level.
The threshold voltage level is used in decoding and will become the trigger
level when the trigger type is set to the selected serial decode bus.
7
Press
Baud
and select the baud rate of the FlexRay signal being probed.
8
Press
Bus
and select the bus type of the FlexRay signal being probed.
It is important to specify the correct bus because this setting affects CRC error
detection.
9
Press
Auto Setup
to perform the following actions:
•
Sets the selected source channel's impedance to 50 Ohms, assuming a
differential active probe that requires a 50 ohm termination is being used .
•
Sets the selected source channel's probe attenuation to 10:1.
•
Sets the trigger level (on the selected source channel) to -300 mV.
•
Turns on trigger Noise Reject.
•
Turns on Serial Decode.
•
Sets the trigger type to FlexRay.
FlexRay Triggering
To set up the oscilloscope to capture a FlexRay signal, see
After you have set up the oscilloscope to capture a FlexRay signal, you can then
set up triggers on frames (see
), errors (see
), or events (see
).
NOTE
To display FlexRay serial decode, see
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...