68
Keysight B2980B User’s Guide, Edition 1
Installation
Installing the Interlock Circuit
Installing the Interlock Circuit
This section is applied to Keysight B2985B/B2987B which supports the interlock
function. The interlock circuit is a simple electric circuit, as shown in
The circuit electrically opens when an access door is opened, and closes when the
door is closed.
The B2985B/B2987B cannot apply high voltages over
21 V when the interlock
terminal is open. To apply high voltage, the B2985B/B2987B interlock terminal
must be connected to the interlock circuit installed in the measurement
environment such as the shielding box. The interlock circuit is important and
necessary to prevent electrical shock when the user touches the output terminal.
Potentially hazardous voltages of up to
1050 V may be present between the
High and Low terminals of the Voltage Source when the interlock circuit is
closed. To prevent electrical shock, do
not
expose the line.
Des tensions potentiellement dangereuses jusqu’à
1050 V risquent d’être
présentes entre les bornes Haute et Basse de la tension de la source lorsque le
circuit de verrouillage est fermé. Afin d’éviter toute décharge électrique,
n’exposez pas ces lignes.
Requirements
•
LED (Keysight part number 1990-0486 or equivalent), 1 ea.
•
Mechanical switch (Keysight N1254A-402 or equivalent), 2 ea.
•
Interlock connector head, 4-pin plug, 1 ea., furnished, Phoenix Contact
1881341 (4 pin) or equivalent
•
Connection wire, needs enough length from the shielding box to the Interlock
connector on the B2985B/B2987B rear panel
Procedure
1. Mount two mechanical switches onto your shielding box, so that the switches
close when the access door is closed, and open when the door is opened. See
for the switch dimensions (typical).
2. Mount an LED onto your shielding box. See
for the LED dimensions
(typical).
Summary of Contents for B2980B Series
Page 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Page 16: ......
Page 24: ......
Page 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Page 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Page 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Page 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...