19
MEMory Subsystem
486
Keysight 8990B Programming Guide
MEMory:TABle:EDIT <numeric_value>,<numeric_value>
This command modifies the offset value of an existing frequency and offset pair in
the FDO table selected in
“MEMory:TABle:SELect <string>”
Syntax
Parameter
Example
Item
Description/Default
Range of values
numeric_value
Numeric values for the frequency and offset pair:
– Minimum values:
– Frequency: 1 kHz
– Offset: 1%
– Maximum values:
– Frequency: 1 THz
– Offset: 150%
The first <numeric_value> represents the frequency while the
second <numeric_value> represents the offset. The frequency
value specified will be truncated to a multiple of 1000.
1 kHz frequency, 1% offset to 1
THz frequency, 150% offset
numeric_value
numeric_value
:EDIT
Space
,
MEM
:TAB
MEM:TAB:EDIT 1000,30
This command changes the offset to 30% for the
1 kHz frequency in the selected FDO table.