8990B Remote Operation
1
Keysight 8990B Programming Guide
45
Using the Operation Complete Commands
The
*OPC?
command allow you to maintain synchronization
between the PC and the 8990B. The
*OPC?
query places an ASCII character 1 into
the 8990B output queue when all pending 8990B commands have completed. If
your program reads this response before continuing program execution, you can
ensure synchronization between one or more instruments and the PC.
The
*OPC
command sets bit 0 (Operation Complete) in the Standard Event Status
register when all pending 8990B operations have completed.
Procedure
1
Send a device clear message to clear the 8990B output buffer.
2
Clear the event registers with the
*CLS
(clear status) command.
3
Enable Operation Complete using the
*ESE 1
command (Standard Event
register).
4
Send the
*OPC?
query and enter the result to assure synchronization.
5
Send your programming command string, and place the
*OPC
command as the
last command.
6
Send the
*STB?
(status byte) query to poll the register. This command does not
clear the Status Byte summary register.
Summary of Commands
For details of each SCPI command available to program the 8990B, refer to later
chapters for more details on each command.
Summary of Contents for 8990B
Page 1: ...Keysight 8990B Peak Power Analyzer Programming Guide ...
Page 16: ...16 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight 8990B Programming Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...