8990B Remote Operation
1
Keysight 8990B Programming Guide
43
Operation Status register
The Operation Status group monitors conditions in the 8990B measurement
process.
The following bits in this register set are used by the 8990B:
Table 1-5
Bit definitions – Operation Status
“Operation Status Register Set”
on page 442 for the
STATus:OPERation
commands.
The condition bits are set and cleared under the following conditions:
Table 1-6
Bit change conditions for the Operation Status register set
Bit
Weight/Decimal
value
Definition
0 to 6
–
Not used
7
128
ARM event occurred
8
256
TRIGger event occurred
9
512
Equivalent-Time Sampling (ETS) operation completed
10 to 15
–
Not used (bit 15 is always 0)
Bit
Definition
EVENts causing bit changes
7
ARM event
occurred
SET:
When the trigger is armed
CLEARED:
When the trigger is unarmed
8
TRIGger event
occurred
SET:
When triggering of a waveform occurs
CLEARED:
When there is no triggering of a waveform
9
ETS operation
completed
SET:
When the ETS operation has completed
CLEARED:
When the ETS operation has not completed