TIMebase Subsystem
7
Keysight 8990B Programming Guide
191
TIMebase:OFFSet <numeric_value>
This command sets the time interval between the trigger event and delay
reference point.
Syntax
Parameter
Example
TIM
Space
?
numeric_value
:OFFS
Item
Description/Default
Range of values
numeric_value
A numeric value in second for the timebase offset:
– Default value: 0 s
– Minimum value:
– –1 s
– –10
μ
s (only applicable when ETS is turned on)
– Maximum value:
– 1 s
– 10
μ
s (only applicable when ETS is turned on)
The range of values will change according to the zoom scale you set in
ACQuire:ZOOM:SCALe <numeric_value>
– –1 s to 1 s
– –10
μ
s to 10
μ
s (only
applicable when ETS is
turned on)
TIM:OFFS 0.05
This command sets the timebase offset to 0.05 s.