4
ACQuire Subsystem
104
Keysight 8990B Programming Guide
ACQuire:CCDF:MARKer:BXPosition <numeric_value>
This command sets the X-axis position of CCDF marker B.
Syntax
Parameter
Example
Remarks
– This command is only applicable for channels 1 and 4. If you send this
command for a channel other than 1 or 4, error 700,“Applicable to channel 1
and 4 only” will occur.
– This command is only applicable in the CCDF acquisition mode. If the
acquisition mode is not set to CCDF, error –221,“Settings conflict;Requires
CCDF mode to be enabled” will occur. Refer to
on page 132 to set the acquisition mode.
– This command is only applicable when the CCDF marker B source is set. If the
source is not set, error –221,“Settings conflict;CCDF marker B source not set”
will occur. Refer to
“ACQuire:CCDF:MARKer:BSOUrce <character_data>”
page 102 to set the CCDF marker B source.
ACQ
:CCDF
Space
?
:BXP
numeric_value
:MARK
Item
Description/Default
numeric_value
A numeric value in dB for the X-axis position
ACQ:CCDF:MARK:BXP 0.5
This command sets the X-axis position of CCDF
marker B to 0.5 dB.