8 Operating and Service Guide 85059-90015
85059G 1.0 mm Connector Gage
Specifications
-
Table 2
Specifications, Characteristics, and Conditions
Comments
Specifications
Specified Limits
Indicator calibration accuracy
0.00012 in (0.003 mm)
From manufacturer specifications
Indicator repeatability
0.000079 in (0.002 mm)
From manufacturer specifications
Typical Characteristics
Typical Limits
Master repeatability
0.00002 in (0.0005 mm)
Functional check of master with gage
Nominal Characteristics
Nominal Value
Net weight
2.6 kg (5.73 lbs)
Battery Life
Approximately 500 hours in
continuous operation
Conditions
Value or Range
Measurement range
-0.005 in (–0.127 mm) to
+0.005 in (+0.127 mm)
Calibrated range for pin depth
measurements in 0.001 mm
increments.
Gage resolution
0.00005 in (0.001 mm)
From manufacturer specifications
Operating environment:
Temperature range
Humidity (at 26°C maximum dry
bulb)
Maximum altitude
+
20° to 26 °C (
+
68 to
+
79 °F)
5 to 80%
4.5 km (15,000 ft)
No temperature fluctuations greater
than 1 °C.
Conditions
Value or Range
Storage and Shipping Environment:
Temperature
Relative Humidity
Maximum Altitude
–10 to 60 °C
Less than 95% at 40 °C